تعارف

The MSA ALTAIR 4XR Multigas Detector is a robust, handheld, portable gas monitor designed for critical safety applications. It is capable of simultaneously detecting up to four gases: Lower Explosive Limit (LEL), Oxygen (O2), Hydrogen Sulfide (H2S), and Carbon Monoxide (CO). Engineered for durability, this device features a tough over-molded rubber armor, is MIL-STD-810G drop protected, and boasts an IP68 rating for dust and water resistance, ensuring reliable performance in demanding environments.

MSA ALTAIR 4XR Multigas Detector front view

شڪل 1: سامهون view of the MSA ALTAIR 4XR Multigas Detector, showing the display and control buttons.

اهم خاصيتون

سيٽ اپ ۽ حساب ڪتاب

Before initial use or after prolonged storage, it is essential to perform a fresh air setup (zero calibration) and a span calibration to ensure accurate readings. Always perform these procedures in a clean, fresh air environment free from contaminants.

گهربل سامان:

Zero Calibration (Fresh Air Setup):

  1. Power on the ALTAIR 4XR detector by pressing and holding the middle button. Allow it to complete its startup sequence and warm-up procedure.
  2. When prompted for a "Fresh Air Setup" (FAS), select "No" to proceed to normal operating mode. This ensures a high-quality zero calibration.
  3. Once in normal operating mode, press and hold the right button until the display shows "ZERO CAL?".
  4. Press the middle button to confirm "Zero Cal". The unit will enter "SENSOR REFRESH" mode. Ensure the detector is in a clean air environment, free from any gas sources or residual gas.
  5. Allow the unit to complete the zero calibration. The display will indicate when it's complete.

اسپين ڪيليبريشن:

  1. Ensure the calibration gas cylinder has the correct gas mixture and is not expired. The standard MSA calibration gas mixture includes: Hydrogen Sulfide (20 ppm), Carbon Monoxide (60 ppm), Pentane Simulant (58% LEL), Oxygen (15.0%), and Nitrogen (Balance).
  2. Attach the 0.25 L/min Regulator to the calibration gas cylinder. Ensure the regulator's valve is open slightly to purge any residual air before fully tightening it onto the cylinder. This prevents moisture from entering the cylinder.
  3. Connect the tubing from the regulator to the calibration adapter.
  4. Attach the calibration adapter to the MSA ALTAIR 4XR detector. The adapter should securely cover the sensor inlets.
  5. From the normal operating mode, press and hold the right button until the display shows "SPAN CAL?".
  6. Press the middle button to confirm "Span Cal". The unit will prompt for gas application.
  7. Immediately turn on the gas flow from the cylinder by opening the regulator valve. The detector will begin to read the gas concentrations.
  8. Monitor the display. The values for each gas (LEL, O2, H2S, CO) should stabilize near the concentrations listed on your calibration gas cylinder. The unit will indicate "SPAN PASS" when calibration is successful.
  9. Once "SPAN PASS" is displayed, turn off the gas flow and remove the calibration adapter and tubing from the detector.
Worker using MSA ALTAIR 4XR Multigas Detector in a confined space

Figure 2: The MSA ALTAIR 4XR Multigas Detector in use, demonstrating its compact size and clear display for field applications.

آپريشن

After successful calibration, the ALTAIR 4XR is ready for use. The large, clear display shows real-time gas concentrations. The unit continuously monitors the environment and will activate audible, visual (LEDs), and vibrating alarms if gas levels exceed predefined thresholds or if a man-down event is detected via MotionAlert.

سار سنڀال

Regular maintenance is crucial for the longevity and accuracy of your MSA ALTAIR 4XR. Keep the device clean and free from debris. Periodically inspect the sensor inlets for blockages. The unit is compatible with MSA GALAXY GX2 Automated Test System for simplified and automated bump testing and calibration, which also helps in managing data and records.

Sensors are designed for long life but may require replacement over time. The device will provide an "end-of-sensor-life" warning. Replacement MSA XCell Sensors are available and can be easily installed in the plug-and-play slots.

مشڪلاتون

If the device fails a bump test or calibration, first ensure that the calibration gas cylinder is not expired and contains the correct mixture. Check all connections for leaks. If sensor readings are erratic or consistently off, the sensor may be poisoned or at the end of its life and require replacement. For persistent issues, refer to the comprehensive troubleshooting guide in the full product manual or contact MSA support.

وضاحتون

تفصيلقدر
برانڊايم ايس اي
اندازO2, H2S, CO
طاقت جو ذريعوبيٽري سان هلندڙ
رنگڪارو
پيداوار جي ماپ1.37"ڊي x 3"ڊي x 4.4"ايڇ
شيءِ جو وزن0.51 پائونڊ (8.1 اونس)
الارمٻڌڻ لائق، بصري
آپريٽنگ نمي95 سيڪڙو
اپر درجه حرارت جي درجه بندي50 درجا Celsius
سينسر جو قسمCatalytic, Electrochemical
شين جو تعداد1
ٺاهيندڙايم ايس اي
حصو نمبر10178557
شيون ماڊل نمبر10178557
تاريخ پهرين دستيابجنوري 25، 2018

وارنٽي ۽ سپورٽ

This product is covered by a 4-year warranty. For more information regarding warranty claims or technical support, please visit the official MSA website or contact their customer service. Ensure to register your product to activate the warranty and receive important updates.

For additional assistance, you may also refer to the comprehensive product manual provided with your device or contact a certified MSA service center.

لاڳاپيل دستاويز - 10178557

اڳيview MSA ALTAIR 4XR Multigas Detector Operating Manual
Comprehensive operating manual for the MSA ALTAIR 4XR Multigas Detector, covering safety regulations, device description, operation, maintenance, technical data, and ordering information.
اڳيview گيس ڊيٽيڪٽرن لاءِ MSA GALAXY GX2 خودڪار ٽيسٽ سسٽم
MSA GALAXY GX2 خودڪار ٽيسٽ سسٽم MSA ALTAIR ۽ ALTAIR PRO سنگل گيس ڊيڪٽرز، ۽ ALTAIR 4XR ۽ ALTAIR 5X ملٽي گيس ڊيڪٽرز لاءِ سادي، ذهين ٽيسٽنگ ۽ ڪيليبريشن پيش ڪري ٿو. ان ۾ استعمال ۾ آسان انٽرفيس، خودڪار ٽيسٽنگ، ۽ جامع فليٽ مئنيجمينٽ صلاحيتون شامل آهن.
اڳيview MSA ALTAIR 5X ۽ 5XIR ملٽي گيس ڊيٽيڪٽر آپريٽنگ مينوئل
MSA ALTAIR 5X ۽ ALTAIR 5XIR ملٽي گيس ڊيٽيڪٽرز لاءِ جامع آپريٽنگ مينوئل، جيڪو صنعتي ماحول ۾ گئس جي ڳولا لاءِ حفاظتي ضابطن، آپريشن، سار سنڀال، ٽيڪنيڪل وضاحتن، ۽ مسئلن جي حل کي ڍڪيندو آهي.
اڳيview MSA ALTAIR پمپ پروب: خطرناڪ ماحول لاءِ پائيدار گيس ڊيٽيڪٽر
MSA ALTAIR پمپ پروب کي ڳولهيو، هڪ مضبوط ۽ قابل اعتماد گئس ڳولڻ وارو اوزار. صنعتي حفاظتي ايپليڪيشنن لاءِ ان جي اهم خاصيتن، ٽيڪنيڪل وضاحتن، منظورين، ۽ آرڊرنگ جي معلومات بابت سکو.
اڳيview MSA Galaxy GX2 Automated Test System Operating Manual
Operating manual for the MSA Galaxy GX2 Automated Test System, detailing installation, setup, operation, maintenance, and troubleshooting for gas detector testing.
اڳيview ايم ايس اي الٽير پرو سنگل گيس ڊيٽيڪٽر آپريٽنگ مينوئل - حفاظت ۽ استعمال جي گائيڊ
هي آپريٽنگ مينوئل MSA ALTAIR Pro سنگل گيس ڊيٽيڪٽر لاءِ تفصيلي هدايتون، حفاظتي هدايتون، ڪيليبريشن طريقا، ۽ مسئلا حل ڪرڻ فراهم ڪري ٿو. سکو ته پنهنجي گيس جي ڳولا واري اوزار کي مؤثر طريقي سان ڪيئن استعمال ڪجي ۽ برقرار رکجي.